Blood Pressure and Vital Signs Monitor, suitable for adults, pediatrics and neonates

Features/Parameters:

Large LED digital display

User-friendly interface

2 hours mains independent operation

Automatic self-test

RS-232 for connectivity and Nurse Call connection

Available in various configurations

Optional: integrated thermal printer

Technical data ARGUS VCM Blood Pressure & Vital Signs Monitor:

Configurations:

- NIBP, PR

- NIBP, PR with printer

- NIBP, PR, SpO2

- NIBP, PR, SpO2 with printer

Blood pressure accuracy:

Meets or exceeds EN 1060-1:1995, EN 1060-3:1997, IEC 60601-2-30,

ANSI/AAMI SP10-1992 + A1:1996

Patient modes:

Selectable modes of neonate, paediatric and adult

Blood pressure range:

Adult:

- Systolic: 30 – 260 mmHg

- Diastolic: 20 – 235 mmHg

- MAP: 20 – 255 mmHg

Pediatric:

- Systolic: 30 – 160 mmHg

- Diastolic: 15 – 130 mmHg

- MAP: 15 – 140 mmHg

Neonatal:

- Systolic: 25 – 120 mmHg

- Diastolic: 10 – 105 mmHg

- MAP: 10 – 110 mmHg

Initial cuff inflation:

- Adult: 160 mmHg

- Pediatric: 120 mmHg

- Neonate: 90 mmHg

Pulse rate:

Pulse rate accuracy:

SpO2 module pulse rate (Nellcor):

- No motion: ± 3 BPM

- During motion: ± 5 BPM

NIBP module pulse rate: ± 3 BPM or 3 %, whichever is greater

Pulse rate range:

- SpO2 module pulse rate range: 20 – 300 BPM

- NIBP module pulse rate range: 30 – 220 BPM

SPO2:

Utilizes Nellcor Oximax Technology

Saturation rate: 0 – 100 %

Probe accuracy: 70 – 100 %; 28 – 42° C

Digit accuracy: Without motion: adult/paediatric: ± 2 %, neonate: ± 3 %

With motion: adult/neonate: ± 3 %, low perfusion: ± 2 %

Power requirements:

North American version: 120 VAC, 60 Hz

International version: 220 - 240 VAC, 50 - 60 Hz

Internal battery: Lead Acid, with internal recharger

Language support:

English, French, German, Italian, Spanish, Portuguese, Dutch, Swedish, Chinese, Japanese,

Polish

Dimensions/weight:

Dimensions (approx.): 130 x 180 x 278 mm

Weight: 2.7 kg