nCompass™ 8100H

The nCompass™ 8100H Series monitor is the perfect choice for busy departments demanding accurate, reliable performance.

The 8100H Series monitor can monitor adult and pediatric patients in virtually every department. The innovative user interface features simplified menus and dedicated function keys for timely setups. The big, bright display with large numerics can easily be seen from across the room.

The 8100H Series monitor mounts quickly and easily to walls and rollstands.

Features and Benefits

  • Easy-to-read 12.1" active color TFT display.
  • Easy to learn, easy to use with simplified menus and dedicated function keys.
  • Mini-Trends can display condensed tabular vital signs data to show recent monitoring activities.
  • Cardio-Respirogram Trend Display.
  • Optional invasive cardiac output interface capability.
  • Compatible with Criticare's 8500 Series Agent Gas Analyzer for anesthetic monitoring.

Technical Specifications

ECG

  • Number of Leads: 3- or 5-Lead
  • Lead View: User-selectable; I, II, III, aVR, aVL, aVF, V (5-Lead); I, II, or III (3-Lead)
  • Gain Selection: x0.5, x1, x2, and x4
  • Frequency Response: Diagnostic: 0.05 to 100 Hz (-3dB);
  • Monitor: 0.5 to 40 Hz (-3dB);
  • ST: 0.05 to 40 Hz (-3dB)
  • Electrosurgery Protection: Yes
  • Defibrillator Protection: Yes
  • Pacer Detection/Rejection: Yes
  • ST Accuracy: ±0.2 mm or ±5% of reading, whichever is greater, over heart range of 30 to 120 bpm
  • ST Range: -6.0 to 6.0 mm

NIBP (Non-Invasive Blood Pressure)

  • Technique: Oscillometric measure upon inflation
  • Measurement Cycle: < 40 sec. typical
  • Auto. Measurement Cycles: 2, 3, 5, 10, 15, 30 min; 1, 2, 4 hr
  • Max. Allowable Cuff Pressure: Adult/Pediatric: 300 mmHg; Neonate: 150 mmHg
  • Range: Adult/Pediatric: 30-300 mmHg; Neonate: 20-150 mmHg
  • Resolution: 1 mmHg
  • Transducer Accuracy: ±2 mmHg or ±2% or reading, whichever is greater,
  • over full range
  • STAT Mode: 5 min of consecutive readings
  • Clinical Accuracy: Complies with AAMI SP-10:2002

Pulse Oximetry

  • Range: 1% to 99%
  • Resolution: 1%
  • Accuracy: 70% to 99% range: ±2%; 50% to 69% range: ±3%;
  • < 50%: unspecified
  • Method: Dual wavelength LED

Heart (Pulse) Rate

  • Source: User selectable: Smart, ECG, Pleth, NIBP, IBP
  • Range: 20 to 300 bpm (ECG, Pleth);
  • 30-240 bpm (NIBP); 30-300 bpm (IBP)
  • Accuracy: ±1 bpm or ±1% (ECG) whichever is greater

Respiration

  • Rate: 6 to 150 bpm (ECG); 2 to 120 bpm (CO2)
  • Resolution: 1 bpm
  • Accuracy: ±2 bpm or ±2%, whichever is greater
  • Source: User selectable: Smart, CO2, ECG

CO2

  • Units: mmHg; %; kPa; Torr.
  • Method: Non-dispersive infrared; autocalibrating.
  • N2O Compensation: Manual (On/Off).
  • Range: 0–99 mmHg; 0–12.5%; 0–12.5 kPa; 0–99Torr.
  • Resolution: 1 mmHg; 0.1%; 0.1 kPa; 1 Torr.
  • Flow, Sample Rate: 200 ml/min.
  • Accuracy: ±2 mmHg; ±0.3 vol. %; ±0.3 kPa; ±2 Torr or ±4% of reading, whichever is greater, at 200 ml/min & respiratory rate of ≤ 120 bpm.
  • Rise Time: 170 ms (10–90%) at 200 ml/min.

IBP (Invasive Blood Pressure)

  • Channels: 1 or 3.
  • Transducer Sites: Selectable: ART, PA, LV, RV, LA, RA, CVP, ICP.
  • Pressure Range: -10 mmHg to 300 mmHg.
  • Transducer: Freq. Response 0–20 Hz; Impedance >300 ohms;
  • Sensitivity 5μV/V/mmHg.
  • Accuracy: ±1 mmHg or ±1%, whichever is greater.

Physical Properties

  • Weight: 12 lbs; 5.4 kg.
  • Size: 12” (H) x 14.5” (W) x 11.5” (D);
  • 30.5 cm (H) x 36.8 cm (W) x 29.2 cm (D)

Power Requirements

  • Voltage: 100 to 240 VAC; 50/60 Hz.
  • Power Consumption: 35 W, typical.
  • Battery: 1 sealed lead acid.
  • Battery Life: 2 hours, typical (3 hours, typical for base unit).
  • Recharge Time: 4 to 6 hours, typical.

Classification

  • Medical Device: Class II Equipment (IIb EU).
  • Electrical Protection: Class I Equipment.
  • Degree of Protection: Type CF, Defibrillator-proof.
  • Protection Against Ingress: IPX1.
  • Safety: EN 60601-1.

Custom Configurations

  • Standard: ECG, Respiration, NIBP, HR, SPO2, Temperature, Arrhythmia and ST analysis, and Nurse Call interface connection.
  • Optional: CO2, 1 or 3 channel IBP, printer.