Size: А А А Interval: AB AB AB Colors: А А А А Images: Включить графику Выключить графику
Regular versionRegular version

FM, SF FLOWMETERS

FM FLOWMETERS

The FM flowmeters are devices for the supply of medical gases with two to six antistatic and graduated tubes and completed with an integrated dosage unit. The body is made of anodized aluminum and the inlet connections for O2, Air and N2O at the bottom of the flowmeter are arranged to be connected to hospital centralized gas plants or to cylinders complete with pressure regulators. The gas feeding unit includes a safety valve automatically blocking the N2O supply when the feeding pressure of O2 drops below the value of approximately 0.8 bar.

The mixtures of O2 and N2O can be supplied to the patient through a dosage unit. An incorporated mixing device proportions the N2O in such a way that a minimum of at least 30% of O2 is always present in the gas mixture. At the outlet of the dosage unit, an anesthesia evaporator can be connected or the gas mixture can be supplied directly to the patient through the gas feeding unit. The O2+ flush button can be used for a quick supply of O2 directly at the outlet of the mixture on the gas feeding unit.

FM 2200

FM 2300

FM 2500

Sizes (LxWxH)

160x145x318 mm

160x145x318 mm

160x145x318 mm

Weight

5.52 Kg

5.57 Kg

5.64 Kg

Dosage range

O2: 0.2 - 15 L/min.
N2O: 0.2 - 12L/min.

O2:
0.1 - 1 L/min.
0.2 - 15 L/min.
N2O:
0.1 - 1 L/min.
0.2 - 12 L/min.

O2: 0.2 - 15 L/min.
Air: 0.2 - 15 L/min.
N2O: 0.2 - 12 L/min.

FM 2800

FM 2900

Sizes (LxWxH)

160x145x318 mm

160x190x318 mm

Weight

5.69 Kg

7.2 Kg

Dosage range

O2:
0.1 - 1 L/min.
0.2 - 15 L/min.
Air:  0.2 - 15 L/min.
N2O::
0.1 - 1 L/min.
0.2 - 12 L/min.

O2:
0.1 - 1 L/min.
0.2 - 15 L/min.
Air:  0.2 - 15 L/min.
N2O:
0.1 - 1 L/min.
0.2 - 12 L/min.
CO2:  0.1 - 2 L/min. (limited at 1 L/min.)

Accuracy

±10% read value or ±0.3 L/min. if greater (± 10% E.O.S. for flow < than 1 L/min.)

Inlet pressures

3.5 - 5 bar ± 20%

N2O blocking device

the supply of N2O is blocked when the O2 pressure drops below 0.8 bar ±0.3 bar

Charge loss level, side under pressure (before the adjusting needle valves)

less than 25 ml/min. in normal pressure conditions (ISO 5358)

Charge loss level, low pressure side(after the adjusting needle valves, vaporizer excluded)

less than 25 ml/min. at 30 cm H2O

Hypoxyguard system

at least 30% -3% O2 is garanteed in the mixture with N2O opening value equal to 0.4 L/min. (with inlet pressures of 3.5 bar)

Mixed gases outlet connection

conical connection 23 mm F ISO DIN 5356/1

Gases inlet connections

hose connector Ø 6 mm

O2 quick dosage
(if foreseen)

when pushed it can supply more than 45 L/min. O2

SF FLOWMETERS

The SF flowmeters are devices for the supply of medical gases particularly suitable in respiratory therapy. They are made of anodized aluminum body, and they are available with one or more graduated tubes and a built-in dosage unit. The inlet connections are at the bottom of the unit and arranged to be connected to the hospitals centralized gas plants or to cylinders complete with pressure regulators. For veterinary use, an anesthesia evaporator can be connected at the outlet of the dosage unit.

SF 1

SF 2

SF 3

SF 4

Sizes (LxWxH)

115x73x294mm

115x114x294mm

115x137x294mm

115x170x294mm

Weight

1.12 Kg

1.61 Kg

2.0 Kg

2.4 Kg

Dosage range

O2: 0.1 - 1 L/min.  • 0.1 - 8 L/min. • 0.2 - 15 L/min.
N2O: 0.1 - 1 L/min. • 0.2 - 12 L/min.
Air:  0.2 - 15 L/min.

Accuracy

±10% read value or ± 0.3 L/min. if greater
(±10% E.O.S. for flow < than 1 L/min.)

Inlet pressures

3.5 - 5 bar ± 20%

Charge loss level, side under pressure (before the adjusting needle valves)

less than 25 ml/min. in normal pressure conditions (ISO 5358)

Charge loss level, low pressure side (after the adjusting needle valves)

less than 25 ml/min. at 30 cm H2O

Mixed gases outlet connection

22 M. - 15 F. ISO DIN 5356-1 or conical connection 23 mm F ISO
DIN 5356/1 (for veterinary use only)

Gases inlet connections

hose connector Ø 6 mm